Comparison Of Manual Versus Automated Blood Pressure Measurement In Intensive Care Unit
Written by Jessica Roggers
OBJECTIVE:
To determine the accuracy of three automatic monitors (arm, wrist, finger) for blood pressure measurement manufactured by Omron compared with a standard mercury sphygmomanometer. PRIMRY END POINT: Difference in the mean blood pressure readings from each monitor; the secondary end point was difference in pulse readings.
DESIGN:
A single-visit, crossover trial tested each device twice on the left arm of each participant; the average of the two readings was recorded. The pulse readings from each monitor were also recorded. ANOVA was used to compare mean blood pressure readings and pulse readings from each device.
RESULTS:
A total of 55 persons (mean age 53 y; 36 women) met inclusion criteria and completed the study. The mean systolic and diastolic readings obtained from the electronic arm unit were comparable to the mercury readings (124.4/78.02 vs. 129.45/77.87 mm Hg, respectively; p > 0.05 for both readings). The mean results obtained from the wrist and finger monitors differed significantly from those of the mercury readings (145.44/89.58 and 113.94/69.07 mm Hg, respectively; p < 0.05 for both monitors compared with control). No difference was measured in the mean pulse readings between the comparisons (p = 0.72). The absolute difference in systolic and diastolic blood pressure readings from control varied the least with the arm monitor.
CONCLUSIONS:
Compared with the mercury sphygmomanometer, the arm monitor was the most accurate in measuring blood pressure. The wrist and finger monitors resulted in statistically significant mean systolic and diastolic differences compared with the mercury sphygmomanometer.
https://www.amazon.com/Santamedical-Adult-Deluxe-Aneroid-Sphygmomanometer/dp/B00ACBKNS6
https://santamedical.com/product/adult-deluxe-aneroid-sphygmomanometer/
Article author
About the Author
To determine the accuracy of three automatic monitors (arm, wrist, finger) for blood pressure measurement manufactured by Omron compared with a standard mercury sphygmomanometer.
Further reading
Further Reading
Article
Fear of Being Healthy?
Why do some people have trouble addressing health concerns?
October 8, 2026
Article
Why General Practice Accreditation Delivers Real Value Beyond Compliance
General practice accreditation is more than a compliance requirement. This article explains how accreditation supports patient safety, quality improvement, practice governance, funding eligibility, and ongoing trust in Australia’s healthcare system.
October 8, 2026
Website
HealthNexo
HealthNexo provides trustworthy, evidence-based health information and the latest health news. We cover disease prevention, diagnosis, nutrition, fitness, and wellness to help you make informed decisions for a healthier life.
October 2, 2026
Article
Dengue Diagnosis: Which Blood Tests Are Used and When?
Dengue diagnosis blood tests include RT-PCR/NAAT, NS1, IgM, IgG, and CBC. Early tests detect the virus, while IgM becomes more useful later. CBC helps monitor the patient but cannot confirm dengue alone.
October 2, 2026